Products
// Electromagnetic Flowmeter |
|
|
|
|
AMC3200DF Series |
|
The AMC3200DF Series converter features in dual frequency excitation. The reduced noise and improved the signal to noise ratio makes it more idea for slurry measurement in various industries such as papermaking and chemistry. The unit features multiple display parameters including flow rate, flow total, diagnostics, programming and troubleshooting. It's equipped with bluetooth, HART and MODBUS communication functions. |
|
|
|
FEATURES |
|
Dual Frequency Excitation Method
Three line LCD display Flow Rate / Totalizer
Accuracy of +/-0.4% value of reading
Bi-directional flow measurement, current and pulse output
Explosion Proof, Ex d (ia) llC T3-T6
Bluetooth / MODBUS / HART communication
Converter can be rotated at Four directions for easy installation
|
|
SPECIFICATION |
|
Excitation: |
Keypad: |
Dual Frequency Excitation |
3 Internal keys for programming |
Display: |
and display control |
3 Lines LCD with illumination |
Data Storage: |
Line 1 |
Operation parameters and totalizer |
8 Digit totalizer for forward flow |
figures are stored by EEPROM for more |
8 Digit totalizer for reverse flow |
than 10 years |
8 Digit totalizer for difference |
Low Flow Cutoff: |
Line 2 |
0-9.9% |
4 Digit flowrate / velocity / % / mA / Hz |
Damping: |
Line 3 |
0-99 Seconds |
Parameter / Query / Clr tot / Query LOG |
Density Setting: |
Alarm, Guidance |
0.001-9.999 g/cm3 |
Totalizer Unit |
Self Diagnosis: |
t, kg, g, lb, Ston, Lton, m3, L, mL, Igal, gal, |
The following error message are |
Mgal, bbl, ft3, a-ft |
displayed if failure occurs |
Time Unit |
-Abnormal sample |
sec, min, hour, day |
-Pulse width |
Measuring Range: |
-Overanged output |
Min. 0-0.15 m/s |
-Internal error |
Max. 0-15 m/s |
-Overflow |
Accuracy: |
-Empty pipe |
+/-0.4% of reading (Velocity>=1.0 m/s) |
Cable Entry: |
+/-0.002 m/s (Velocity: 0.3...1.0 m/s) |
Standard: M20 Option: 1/2" NPTF |
Repeatability: |
Protection Class: |
+/-0.2% of reading |
IP67, Explosion Proof, Ex d (ia) llC T3-T6 |
Current Output: |
Ambient Temperature: |
4-20 mA (Isolated) |
-25~65 oC |
Load |
Material: |
Max. 500  |
Aluminum Alloy |
Pulse Output: |
Power Supply: |
Open collector |
90-260 VAC, 50/60 Hz |
Rating |
24 VDC +/-10% (Option) |
3-30 VDC, 50 mA Max. |
Power Consumption: |
Pulse Rate: |
6-8 W (Dependent on sensor size) |
1)Scale pulse (mL/Pusle, L/Pulse, Other) |
Weight: |
2)Frequency: 0-5000 Hz |
4.5 kg |
Pulse Width: |
Mounting: |
50% of band width |
Bracket on 2" pipe |
Alarm: |
AMF Collocation Type: |
High / Low with open collector output |
AMF900, AMF601, AMF500 |
Rating |
Sensor Size: |
3-30 VDC, 50 mA Max. |
DN50-DN450 |
Communication: |
|
Bluetooth (Android) Standard |
|
RS485 (MODBUS Protocol) Standard |
|
HART signal (Compatible) Optional |
| |
|
|
|
 |
|
|